OPERATING MANUAL
REFLEKS™ HP/MD SERIES
HP/MD1015 – HP/MD1025 – HP/MD1035

 

POWER UP/POWER DOWN

To power up – Pretty simple.  Plug the battery into the HP/MD jack marked "Battery Power"

To power down – Even more simple.  Unplug the battery form the HP/MD jack marked "Battery Power". You may stow the battery’s male connector in the battery’s female charging jack for convenience.

TAKING INVENTORY READINGS

If you have previously completed the HP/MD tank programming section:

1.  Power the unit up by plugging the battery pack into the power receptacle on the HP/MD.

2.  Take note of the battery voltage displayed.  The unit will warn you if the voltage is to high or getting too low.  Press the "M" key to advance.  The unit will go through a short initialization sequence and display Tank?

3.  Plug the unit’s fiber optic cable connector into the keyed optical receptacle in the enclosure box.  If your sensor system is equipped with the optional MDV shutter device, don’t forget to open the shutter.

4.  At the Tank? prompt, enter a valid tank ID# number that has been previously programmed into the unit.  If you have not yet programmed tank(s) into the unit, please read the Programming Tanks section at this time.

5.  Press the "M" key to begin the measuring process.  The display will first indicate Staging, then Measuring, four (4) audible beeps, then Finishing.  After the four beeps and at the Finishing display message, you could actually disconnect the fiber connection and be on your way.  The unit is actually just doing some internal calculations before it displays the answer.

NOTE: If sufficient light is not received within 7 seconds, a "low signal" error message is displayed and the measuring process must be restarted.

6.  The HP/MD defaults to displaying an Outage reading.  After Finishing, the REFLEKS HP/MD will display Out = XXXX and the abbreviated measuring units (inches, meters, feet&inches etc.).  Press the "M" key again.  The display will read Save to LTM = 1(long-term memory). Press "1" if you wish to save the reading in LTM or "M" if not.  The HP/MD has the ability to store a series of tank inventories (up to 50), that can be recalled sequentially.  At the end of each inventory, the REFLEKS HP/MD will ask if you wish to store the inventory reading just taken.  See the section entitled "Recalling Sequentially Stored Tanks" regarding recall of this information.   This function does not affect the features offered under "Memory Recall".   The most previous inventory reading is always stored in memory recall without regard to your LTM choice.